PSTatistic Subsystem 9
N1911A/1912A P-Series Power Meters Programming Guide 339
PSTatistic[1]|2:CCDF:CONTinuous <boolean>
This command is used to turn on or off the CCDF Continuous Refresh mode for
Channel A or Channel B.
Syntax
Example
Reset Condition
On reset, the CCDF Continuous Refresh mode will be turned on.
This command is only applicable when P-Series sensor is present and free-run
acquisition mode is selected. If P-Series sensor is used by the adjacent channel
in a dual channel setup, the same setting will be applied for free-run acquisition
mode.
PST
2
:CCDF
:CONT
1
0|OFF
1|ON
Space
?
PST1:CCDF:CONT ON This command turns on the CCDF
Continuous Refresh mode for Channel A.
PST2:CCDF:CONT OFF This command turns off the CCDF
Continuous Refresh mode for Channel B and
the Single Refresh mode is on.