10 SENSe Subsystem
388 N1911A/1912A P-Series Power Meters Programming Guide
[SENSe[1]]|SENSe2:BUFFer:COUNt <numeric_value>
This command sets the buffer size for average or peak measurement. It must be
used in conjunction with external trigger.
It can only be set when frequency sweep is disabled (FREQ:STEP 0). Otherwise,
this parameter will be automatically overwritten by frequency sweep step. If trace
display is turned on, the measurement window will be restored to single numeric
or analog depends on the number of measurement channel.
Syntax
Parameters
NOTE
This command is only applicable when used with 8480, N8480, E4410, E9300,
E9320 or N1920 power sensors (average or peak).
Item Description Range of Values
numeric_value A numeric value for buffer size. 1 to 2048
:COUN
Space
?
numeric_value
SENS
1
SENS2
FREQ
: