SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 389
Example
Query
[SENSe[1]]|SENSe2:BUFFer:COUNt?
This query is used to retrieve the average or peak measurement buffer size.
Query Example
On Reset
On *RST, the value is set to 1.
Error Messages
If no sensor or wrong sensor is connected to the channel, error –241,
“Hardware missing” occurs.
If E4410, N8480, E9300, E9320 or N1920 sensor is connected but acquisition
mode is in free run, error –221, “Setting conflict. Invalid acquisition mode”
occurs.
If frequency sweep step is non-zero, error –221, “Settings conflict. Frequency
sweep enabled. Buffer count overidden” occurs.
If parameter is set lower than 1, error –222 “Data out of range; value clipped to
lower limit” occurs.
If parameter is set higher than 2048, error –222 “Data out of range; value
clipped to upper limit” occurs.
BUFF:COUN 100 This command sets the average or peak
measurement buffer size to 100 for Channel A.
BUFF:COUN? This query returns the average or peak
measurement buffer size for Channel A.