SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 389
Example
Query
[SENSe[1]]|SENSe2:BUFFer:COUNt?
This query is used to retrieve the average or peak measurement buffer size.
Query Example
On Reset
On *RST, the value is set to 1.
Error Messages
– If no sensor or wrong sensor is connected to the channel, error –241,
“Hardware missing” occurs.
– If E4410, N8480, E9300, E9320 or N1920 sensor is connected but acquisition
mode is in free run, error –221, “Setting conflict. Invalid acquisition mode”
occurs.
– If frequency sweep step is non-zero, error –221, “Settings conflict. Frequency
sweep enabled. Buffer count overidden” occurs.
– If parameter is set lower than 1, error –222 “Data out of range; value clipped to
lower limit” occurs.
– If parameter is set higher than 2048, error –222 “Data out of range; value
clipped to upper limit” occurs.
BUFF:COUN 100 This command sets the average or peak
measurement buffer size to 100 for Channel A.
BUFF:COUN? This query returns the average or peak
measurement buffer size for Channel A.