14 TRIGger Subsystem
660 N1911A/1912A P-Series Power Meters Programming Guide
TRIGger[:SEQuence]:DELay <numeric_value>
This command sets the delay between the recognition of a trigger event and the
start of a measurement.
Syntax
Parameters
Item Description/Default Range of Values
numeric_value The delay between the recognition of a trigger event and
the start of the measurement.
DEF: the default value is 0 seconds
Units are resolved to 1.25 ns.
–1 to 1 second
DEF
:DEL
Space
?
numeric_value
DEF
TRIG
:SEQ
NOTE
Trigger delay is not applicable when the power meter is set to power sweep
mode or frequency sweep mode.