MEMory Subsystem 7
N1911A/1912A P-Series Power Meters Programming Guide 291
Table 7-3 Frequency and Calibration/Offset Factor List
[a] Default is a sensor calibration table in which the reference calibration factor and calibration factors are 100%. This sensor calibration table
can be used during the performance testing of the power meter.
For sensosr calibration tables (8480 Series sensors and N8480 Series sensors with
Option CFT only), the number of frequency points must be one less than the
number of calibration factor points. This is verified when the sensor calibration
table is selected using SENSe:CORRection:CSET:SELect <string>.
Ensure that the frequency points you use cover the frequency range of the signals
that you want to measure. If you measure a signal with a frequency outside the
frequency range defined in the table, then the power meter uses the highest or
lowest point in the table to calculate the calibration factor/offset.
Depending on available memory, the power meter is capable of storing 20 sensor
calibration tables and 10 frequency dependent offset tables, each containing
80 points.
Table Power Sensor Table Name
0None
DEFAULT
[a]
1 8481A 8481A
2 8482A, 8482B, 8482H 8482A
3 8483A 8483A
4 8481D 8481D
5 8485A 8485A
6 R8486A R8486A
7 Q8486A Q8486A
8 R8486D R8486D
9 8487A 8487A