EasyManua.ls Logo

Keysight N1911A

Keysight N1911A
771 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OUTPut Subsystem 8
N1911A/1912A P-Series Power Meters Programming Guide 307
OUTPut:RECorder[1]|2:LIMit:LOWer <numeric_value>
This command sets the minimum scaling value for the specified recorder output.
The units used are dependent on the units currently set for the CALC block
specified in OUTPut:RECorder[1]|2:FEED <data_handle>.
Syntax
Parameters
Example
Reset Condition
On reset, the minimum scaling value is set to –150 dBm.
Item Description/Default Range of Values
numeric_value A numeric value for the minimum scaling value. The units
used—dBm, W or %—are dependent on the units currently
set for the CALC block specified in
OUTPut:RECorder[1]|2:FEED <data_handle>.
–150 to +230 dBm
1 aW to 100 XW
0 % to 999 %
OUTP:REC:LIM:LOW –90 This command sets the minimum scaling
value to –90.
OUTP
:LIM
:REC
Space
?
:LOW
numeric_value
1
2

Table of Contents

Other manuals for Keysight N1911A

Related product manuals