3 CALCulate Subsystem
188 N1911A/1912A P-Series Power Meters Programming Guide
CALCulate[1]|2|3|4:LIMit:CLEar:AUTo <boolean>|ONCE
This command controls when the FCO (fail counter) is cleared of any limit failures.
The FCO is used to determine the results returned by the
CALCulate[1]|2|3|4:LIMit:FAIL? query.
–If ON is specified, the FCO is set to 0 each time a measurement is:
Initiated using INITiate[:IMMediate]
Initiated using INITiate:CONTinuous ON
–Measured using MEASure?
Read using READ?
–If OFF is specified, the FCO is not cleared by the above commands.
–If ONCE is specified, the FCO is cleared only after the first initialization then
starts accumulating any limit failures.
Syntax
0|OFF
1|ON
Space
?
:LIM
:CLE
:AUTO
ONCE