CALibration Subsystem 4
N1911A/1912A P-Series Power Meters Programming Guide 221
CALibration[1]|2:RCALibration <boolean>
This command enables and disables the zero/cal lockout facility. With the lockout
facility enabled the power meter is stopped from making measurements until the
connected sensor has been zeroed and calibrated.
Syntax
Example
Reset Condition
On reset, the state of the zero/cal lockout is unaffected.
Query
CALibration[1]|2:RCALibration?
The query enters a 1 or 0 into the output buffer indicating whether zero/cal
lockout is enabled or disabled.
1 is returned if zero/cal lockout is enabled
0 is returned if zero/cal lockout is disabled
CAL1:RCAL 1 This command enables the zero/cal lockout
facility on Channel A.
:RCAL
0|OFF
1|ON
Space
?
CAL
1
2