14 TRIGger Subsystem
646 N1911A/1912A P-Series Power Meters Programming Guide
INITiate[1]|2[:IMMediate]
This command sets the power meter in the wait for trigger state. When a trigger is
received, the measurement is taken and the result placed in the power meter
memory. If TRIGger:SOURce is set to IMMediate the measurement begins as
soon as INITiate:IMMediate is executed.
Use FETCh? to transfer a measurement from memory to the output buffer. Refer
to “FETCh[1]|2|3|4 Queries” on page 132 for further details.
Syntax
Example
Error Messages
If the power meter is not in the idle state or INITiate:CONTinuous is ON, error
–213, “INIT ignored” occurs.
NOTE
This command performs the same function as
INITiate:[IMMediate]:SEQuence[1]|2.
INIT2:IMM This command places Channel B in the wait
for trigger state.