TRIGger Subsystem 14
N1911A/1912A P-Series Power Meters Programming Guide 645
Reset Condition
On reset (*RST), this command is set to OFF.
On preset (SYSTem:PRESet) and instrument power-up, when entering local
mode, if TRIGger[:SEQuence[1]|2]:SOURce is set to INT[[1]|2] or EXT,
INITiate:CONTinuous is not changed. For other trigger sources,
INITiate:CONTinuous is set to ON.
Query
INITiate[1]|2:CONTinuous?
The query enters a 1 or 0 into the output buffer.
1 is returned when there is continuous triggering
0 is returned when there is only a single trigger
Query Example
INIT2:CONT? This command queries whether Channel B is
set for single or continuous triggering.