DISPlay Subsystem 5
N1911A/1912A P-Series Power Meters Programming Guide 243
Parameters
Example
Reset Condition
On reset, the N1911A power meter upper window is DIGital and the lower
window ANALog. For the N1912A power meter, the defaults for the upper and
lower windows are DIGital.
Query
DISPlay:[WINDow[1]|2]:FORMat?
The query returns the current format of the selected window.
Item Description/Default Range of Values
character_data Sets the window format:
– DIGital: sets the window display to digital. This setting is
the same as SNUMeric.
– ANALog: sets the window display to analog using the
currently SELected measurement.
– SNUMeric: sets the window display to single numeric. The
currently SELected measurement is displayed. This setting is
the same as DIGital.
– DNUMeric: sets the window display to dual numeric.
– TRACe: trace display using the currently SELected
measurement. Used to determine the channel from which
the trace is taken.
– CTRAce: sets the display to expanded CCDF window.
– CTABle: sets the window display to CCDF table.
DIGital
ANALog
SNUMeric
DNUMeric
TRACe
CTRAce
CTABle
DISP:WIND2:FORM DIG This command sets the lower window to a
digital display.

Table of Contents

Other manuals for Keysight N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N1911A and is the answer not in the manual?

Keysight N1911A Specifications

General IconGeneral
BrandKeysight
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals