SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 393
[SENSe[1]]|SENSe2:CORRection:CFACtor|GAIN[1][:INPut]
[:MAGNitude] <numeric_value>
This command is used to enter a gain correction value for the calibration factor.
The power meter corrects every measurement by this factor to compensate for the
gain.
Either CFACtor and GAIN1 can be used in the command—both have an identical
result. Using GAIN1 complies with the SCPI standard, whereas CFACtor does
not—this may make your program easier to understand.
Syntax
SENS
1
CORR
SENS2
:
:INP
:MAGN
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space
:GAIN
1
:CFAC