14 TRIGger Subsystem
654 N1911A/1912A P-Series Power Meters Programming Guide
TRIGger[1]|2:DELay:AUTO <boolean>
This command is used to determine whether or not there is a settling-time delay
before a measurement is made.
When this command is set to:
– ON, the power meter inserts a settling-time delay before taking the requested
measurement. This settling time allows the internal digital filter to be updated
with new values to produce valid, accurate measurement results. The trigger
with delay command allows settling time for the internal amplifiers and filters.
It does not allow time for power sensor delay.
In cases of large power changes, the delay may not be sufficient for complete
settling. Accurate readings can be assured by taking two successive
measurements for comparison.
– OFF, the power meter makes the measurement immediately a trigger is
received.
TRIGger[1]|2:DELay:AUTO is ignored if TRIGger[1]|2[:IMMediate] is used
to trigger the power meter.
Syntax
TRIG
:AUTO
:DEL 0|OFF
1|ON
Space
?
1
2
NOTE
Trigger delay is not applicable when the power meter is set to power sweep
mode or frequency sweep mode.

Table of Contents

Other manuals for Keysight N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N1911A and is the answer not in the manual?

Keysight N1911A Specifications

General IconGeneral
BrandKeysight
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals