SERVice Subsystem 15
N1911A/1912A P-Series Power Meters Programming Guide 695
SERVice:BIST:TBASe:STATe <boolean>
This command sends a 10 MHz time base signal to the rear panel trig out for
testing purposes.
If the command is set to:
ON, the 10 MHz time base signal is sent to the rear panel trigger out connector.
OFF, the 10 MHz time base signal is disabled.
Syntax
Example
NOTE
This command overrides the OUTPut:TRIGger[:STATe] command.
For example, if OUTPut:TRIGger[:STATe] is ON and the command
SERV:BIST:TBAS ON is sent, this command overrides the Trigger state and sets
it to OFF. However, the 10 MHz remains out the Trig out port.
If the SERV:BIST:TBAS ON has been sent, the 10 MHz is on and the
OUTPut:TRIGger[:STATe] is then toggled to ON, the channel trigger is now
routed to the Trig out overriding the service command turning the 10 MHz to off.
SERV:BIST:TBAS:STAT OFF This command disables the signal.