10 SENSe Subsystem
438 N1911A/1912A P-Series Power Meters Programming Guide
Example
Query
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:AUTO:REF1|REF2?
The query returns the current setting of Auto Gating Marker Reference 1 or 2 for
the selected gate in numerical value.
Query Example
Error Messages
– If no sensor or wrong sensor is connected to the channel, error –241
"Hardware missing" occurs.
– If channel is not in trigger acquisition mode, error –221 "Settings conflict"
occurs.
– If limits of the values keyed in are exceeded, error –222 "Data out of range;
upper (or lower) limit exceeded; no change" occurs.
SENS1:SWE1:AUTO:REF1 10.0 This command sets the Channel A Auto
Gating Marker Reference 1 to 10 % for
Gate 1.
SENS2:SWE2:AUTO:REF2 40.0 This command sets the Channel B Auto
Gating Marker Reference 2 to 40% for
Gate 2.
SENS1:SWE2:AUTO:REF1? The query returns the current Gate 2
Reference 1 value of Auto Gating Marker.