MEMory Subsystem 7
N1911A/1912A P-Series Power Meters Programming Guide 275
Table 7-2 8480 Series Power Sensor Tables
[a] Default is a sensor calibration table in which the reference calibration factor and calibration factors are 100%. This sensor calibration table
can be used during the performance testing of the power meter
There are also ten sensor calibration tables named CUSTOM_0 through
CUSTOM_9 and ten frequency dependent offset tables named CUSTOM_A
through CUSTOM_J which do not contain any data when the power meter is
shipped from the factory.
Syntax
Example
Table Power Sensor Table Name
0None
DEFAULT
[a]
1 8481A 8481A
2 8482A, 8482B, 8482H 8482A
3 8483A 8483A
4 8481D 8481D
5 8485A 8485A
6 R8486A R8486A
7 Q8486A Q8486A
8 R8486D R8486D
9 8487A 8487A
MEM:CAT:TABL? This command queries the list of stored
tables.
MEM
:CAT
:TABL ?