MEMory Subsystem 7
N1911A/1912A P-Series Power Meters Programming Guide 271
MEMory:CATalog[:ALL]?
This command lists stored sensor calibration tables (8480 Series sensors and
N8480 Series Sensors with Option CFT only), frequency dependent offset tables
and save/recall registers.
The power meter returns the data in the form of two numeric parameters and as
many strings as there are stored tables and save/recall registers:
<numeric_value>,<numeric_value>{,<string>}
– The first numeric parameter indicates the amount of memory, in bytes, used
for the storage of tables and registers.
– The second numeric parameter indicates the memory, in bytes, available for
the storage of tables and registers.
– Each string parameter returned indicates the name, type and size of a stored
table or save/recall register:
– <string>, <type>, <size>
– <string> indicates the name of the table or save/recall register.
– <type> indicates TABL for sensor calibration and frequency dependent
offset tables, or STAT for a save/recall register.
– <size> indicates the size of the table or save/recall register in bytes.
A sample of a response may look like the following:
1178,26230,"DEFAULT,TABL,14","8481A,TABL,116",
"8482A,TABL,74",..........."State0,STAT,1619",
"State1,STAT,1619","State2,STAT,1619" ...........
The power meter is shipped with a set of predefined sensor calibration tables. The
data in these sensor calibration tables is based on statistical averages for a range
of Keysight Technologies power sensors. These tables can be edited. The
predefined data is listed in your user’s guide. These power sensors and table
numbers are listed in Table 7-1.
Predefined sensor calibration table is not applicable for N8480 Series power
sensors with Option CFT. Therefore you are required to create a new sensor
calibration table for the sensors when a sensor calibration table is needed.