7 MEMory Subsystem
268 N1911A/1912A P-Series Power Meters Programming Guide
MEMory Subsystem
The MEMory command subsystem is used to:
– Edit and review sensor calibration tables (8480 Series sensors and N8480
Series sensors with Option CFT only)
– Store sensor calibration tables (8480 Series sensors and N8480 Series sensors
with Option CFT only)
– Edit and review sensor frequency dependent offset tables
– Store sensor frequency dependent offset tables
– Edit and review sensor save/recall registers
Stored tables remain in the power meter’s memory during power down. The
power meter is capable of storing 20 sensor calibration tables and 10 frequency
dependent offset tables of 80 frequency points each.
The MEMory subsystem is not used for E-Series, N8480 Series (excluding Option
CFT) and P-Series power sensors calibration tables. These are automatically
downloaded to the power meter and cannot be reviewed or edited.
Keyword Parameter Form Notes Page
MEMory
:CATalog
[:ALL]? [query only] page 271
:STATe? [query only] page 273
:TABLe? [query only] page 274
:CLEar
[:NAME] <character_data> [no query],
[non-SCPI]
page 277
:TABLe [no query] page 279
:FREE
[:ALL]? [query only] page 281