MEMory Subsystem 7
N1911A/1912A P-Series Power Meters Programming Guide 299
MEMory:TABLe:MOVE <character_data>,<character_data>
This command is used to rename a sensor calibration table (8480 Series sensors
and N8480 Series sensors with Option CFT only) or a frequency dependent offset
table.
Syntax
Parameters
Example
Item Description/Default Range of Values
character_data
(1st parameter)
Contains the existing table name. existing table name
character_data
(2nd parameter)
Details the new table name. A maximum of 12 characters
can be used.
A to Z (uppercase)
a to z (lowercase)
0 - 9
_ (underscore)
MEM:TABL:MOVE "tab1","tab1a" This command renames a table named tab1
to tab1a.
MEM
:TABL
:MOVE
,
Space
character_data
character_data

Table of Contents