CALibration Subsystem 4
N1911A/1912A P-Series Power Meters Programming Guide 227
CALibration[1]|2:ZERO:NORMal:AUTO <boolean>
This command provides a quick way of zeroing the NORMAL path of an E9320
Series sensor. The average only path is unaffected. This command can only be
used to zero an E9320 Series sensor and a P-Series sensor.
The command causes the power meter to perform its zeroing routine, on the
specified channel, when ONCE is selected. This adjusts the power meter for a zero
power reading with no power supplied to the power sensor.
The 0|OFF parameter is only required for the query response and is ignored in the
command. If 1|ON is selected on an E9320 Series sensor, it causes the error –224,
“Illegal parameter value” to occur.
Except when using a P-Series sensor, this command assumes that the E9320
Series sensor is not connected to a power source.
Syntax
Example
The P-Series sensor only has a NORMAL path. Hence, the reason this E9320
Series sensor command is allowed to function.
CAL2:ZERO:NORM:AUTO ONCE This command causes the power meter to
perform a zeroing routine on Channel B.
:AUTO
:NORM
:ZERO
CAL
1
2
0|OFF
Space
?
ONCE