12 SYSTem Subsystem
504 N1911A/1912A P-Series Power Meters Programming Guide
SYSTem:COMMunicate:LAN:DGATeway <character_data>
This command sets the LAN IP router/gateway address for the power meter.
Syntax
Parameters
Example
Query
SYSTem:COMMunicate:LAN:DGAT?
The query returns the current setting of the LAN gateway address.
SYST
:COMM
:LAN :DGAT
Space
?
character_data
Item Description Range of Values
character_data Numeric character values for the address. Up to 15
characters, formatted as follows: A.B.C.D
where A, B, C, D = 0 to 255
0 to 255 (no embedded spaces)
SYST:COMM:LAN:DGAT
‘130.2.6.200’
This command sets the gateway address to
130.2.6.200.