SERVice Subsystem 15
N1911A/1912A P-Series Power Meters Programming Guide 713
SERVice:SENSor[1]|2:PCALfactor <cal_factor_data>
This command writes calibration factor data to, or reads calibration factor data
from, the currently connected sensor. The whole calibration factor block must be
written at once as a checksum is generated. The new block must not be larger
than the existing block.
This command applies to E9320 Series sensors for peak path data only. For E4410
Series, E9300 Series and E9320 Series sensors, average path data, refer to
“SERVice:SENSor[1]|2:CALFactor <cal_factor_data>” on page 705.
Syntax
Parameters
Query
SERVice:SENSor[1]|2:PCALfactor?
The query returns the current peak path calibration factor block.
Item Description/Default
cal_factor_data A binary data block. Refer to Appendix A, “Calibration Factor Block Layout,” on page 759 for
further information.
SERV
1
2
:PCAL
:SENS
cal_factor_data
?
Space