10 SENSe Subsystem
440 N1911A/1912A P-Series Power Meters Programming Guide
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:OFFSet:TIME <numeric_value>
This command sets the delay between the delayed trigger point and the start of
the time-gated period (the offset time) for a P-Series sensor or a E9320 sensors
set to NORMal mode. To set an E9320 sensor to NORMal mode, refer to the
command “[SENSe[1]]|SENSe2:DETector:FUNCtion <character_data>” on
page 414.
Syntax
Parameters
Item Description/Default Range of Values
numeric_value The delay between the trigger point and the start of the
time-gated period.
DEF: the default value is 0 seconds
Units are resolved to 1 ns.
–1 to 1 second
DEF
SENS
1
:TIME
SENS2
:
:OFFS
numeric_value
DEF
Space
SWE
?
1
2
3
4