SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 441
Example
Reset Condition
On reset, the value is set to 0 seconds.
Query
SENSe[1]]|SENSe2:SWEep[1]|2|3|4:OFFSet:TIME?
The query returns the current delay between the trigger point and the start of the
time-gated period.
Query Example
Error Messages
If the command is used when a sensor other than a P-Series or E9320 power
sensor is connected, error –241, “Hardware missing” occurs.
If the command is used when an E9320 sensor is connected and set to AVERage
mode rather than NORMal mode, the error –221, “Settings Conflict” occurs.
SENS2:SWE3:OFFS:TIME 0.001 This command sets the delay to 0.001
seconds.
SENS2:SWE2:OFFS:TIME? The query returns the current delay between
the trigger point and the start of the
time-gated period for Channel B and gate 2.