10 SENSe Subsystem
420 N1911A/1912A P-Series Power Meters Programming Guide
Example
Query
[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed]:STARt?
This query is used to retrieve start frequency (average or peak). Frequency
returned is in Hz.
Query Example
On Reset
On *RST, the value is set to 50 MHz.
Error Messages
If no sensor or wrong sensor is connected to the channel, error –241,
“Hardware missing” occurs.
If parameter set is lower than 1 kHz, error –222, “Data out of range; value
clipped to lower limit” occurs.
If parameter set is higher than 1000 GHz, error –222, “Data out of range; value
clipped to upper limit” occurs.
FREQ:STAR 1 MHz This command sets frequency sweep to start
at 1 MHz for Channel A.
FREQ:STAR? This query returns the start frequency of
frequency sweep in Hz for Channel A.