13 TRACe Subsystem
634 N1911A/1912A P-Series Power Meters Programming Guide
TRACe[1]|2:STATe <boolean>
This command enables or disables trace capture for the specified channel.
Syntax
Example
Reset Condition
On reset trace capture is set to OFF.
Query
TRACe[1]|2:STATe?
The query command enters a 1 or 0 into the output buffer indicating whether or
not trace capture is enabled or disabled.
– 1 is returned when trace capture is enabled
– 0 is returned when trace capture is disabled
This command does not allow the setting set to ON when either measurement
channel (for dual channel) is configured to initiate trigger buffering.
TRAC
0|OFF
1|ON
Space
?
:STAT
1
2
TRAC2:STAT 1 This command enables trace capture for
Channel B.