14 TRIGger Subsystem
664 N1911A/1912A P-Series Power Meters Programming Guide
TRIGger[:SEQuence]:HYSTeresis <numeric_value>
This command sets:
How far a signal must fall below TRIG:LEVel before a rising edge can be
detected.
How far a signal must rise above TRIG:LEVel before a falling edge can be
detected.
Syntax
Parameters
Item Description/Default Range of Values
numeric_value How far a signal must fall/rise before a rising or falling edge can
be detected.
DEF: the default value is 0 dB
Units are resolved to 0.05 dB.
0 to 3 dB
DEF
:HYST
Space
numeric_value
?
DEF
TRIG
:SEQ

Table of Contents

Other manuals for Keysight N1911A