17 IEEE 488.2 Command Reference
738 N1911A/1912A P-Series Power Meters Programming Guide
*ESE <NRf>
The *ESE (Event Status Enable) <NRf> command sets the Standard Event Status
Enable Register. This register contains a mask value for the bits to be enabled in
the Standard Event Status Register. A 1 in the Enable Register enables the
corresponding bit in the Status Register, a 0 disables the bit. The parameter
value, when rounded to an integer and expressed in base 2, represents the bit
values of the Standard Event Status Enable Register. Table 17-1 shows the
contents of this register.
Table 17-1 *ESE Mapping
Syntax
Bit Weight Meaning
0 1 Operation Complete
1 2 Request Control (not used)
24 Query Error
3 8 Device Dependent Error
4 16 Execution Error
5 32 Command Error
6 64 Not used
7 128 Power On
*ESE