14 TRIGger Subsystem
666 N1911A/1912A P-Series Power Meters Programming Guide
TRIGger[:SEQuence]:LEVel <numeric_value>
This command sets the power level at which a trigger event is recognized.
Syntax
Parameters
[a] If a channel offset has been previously set, a higher numeric value is permitted. See“Setting Offsets” on page 61 for more information.
Item Description/Default
Range of Values
[a]
numeric_value The power level at which a trigger event is recognized.
– DEF: the default value is 0 dBm
Units are resolved to 0.1 dBm.
–40 to 20 dBm
DEF
:LEV
Space
numeric_value
?
DEF
TRIG
:SEQ