4 CALibration Subsystem
222 N1911A/1912A P-Series Power Meters Programming Guide
Query Example
Error Messages
When CAL[1]|2:RCAL is ON and the sensor currently connected to the appropriate
channel (A or B) has not been zeroed and calibrated, then any SCPI command
which would normally return a measurement result (for example, FETC?, READ?,
MEAS? etc) does not return a result and generates the error –230, “Data corrupt or
stale; Please zero and Cal.”
After the sensor has been zeroed and calibrated the return measurement results
commands function normally.
CAL1:RCAL? This command queries whether or not the
zero/cal lockout facility is enabled for
Channel A.