Power Meter Remote Operation 1
N1911A/1912A P-Series Power Meters Programming Guide 65
CALCulate[1|2]:LIMit:FAIL? and the
CALCulate[1|2]:LIMit:FCOunt? for window limits
2 Use the STATus command subsystem
Using SENSe and CALCulate
Using SENSe to check the channel limit failures in Figure 1-5 would return the
following results:
Use the equivalent CALCulate commands for checking window limit failures.
SENSe:LIMit:FAIL? Returns 1 if there has been 1 or
more limit failures or 0 if there have
been no limit failures. In this case 1
is returned.
SENSe:LIMit:FCOunt? Returns the total number of limit
failures, in this case 2.