SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 423
Example
Query
[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed]:STEP?
This query is used to retrieve the number of steps in average or peak frequency
sweep.
Query Example
On Reset
On *RST, the value is set to 0.
Error Messages
– If no sensor or wrong sensor is connected to the channel, error –241,
“Hardware missing” occurs.
– If E4410, N8480 (excluding Option CFT), E9300, E9320 or N1920 sensor is
connected but acquisition mode is in free run, error –221, “Setting conflict.
Invalid acquisition mode” occurs.
– If parameter set is lower than 0, error –222, “Data out of range; value clipped
to lower limit” occurs.
– If parameter set is higher than 2048, error –222, “Data out of range; value
clipped to upper limit” occurs.
FREQ:STEP 10 This command sets frequency sweep with 10
steps for Channel A.
FREQ:STEP? This query returns the number of steps in
frequency sweep for Channel A.