SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 423
Example
Query
[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed]:STEP?
This query is used to retrieve the number of steps in average or peak frequency
sweep.
Query Example
On Reset
On *RST, the value is set to 0.
Error Messages
If no sensor or wrong sensor is connected to the channel, error –241,
“Hardware missing” occurs.
If E4410, N8480 (excluding Option CFT), E9300, E9320 or N1920 sensor is
connected but acquisition mode is in free run, error –221, “Setting conflict.
Invalid acquisition mode” occurs.
If parameter set is lower than 0, error –222, “Data out of range; value clipped
to lower limit” occurs.
If parameter set is higher than 2048, error –222, “Data out of range; value
clipped to upper limit” occurs.
FREQ:STEP 10 This command sets frequency sweep with 10
steps for Channel A.
FREQ:STEP? This query returns the number of steps in
frequency sweep for Channel A.