10 SENSe Subsystem
422 N1911A/1912A P-Series Power Meters Programming Guide
Syntax
Parameters
NOTE
This command is only applicable when used with E4410, N8480 (excluding
Option CFT), E9300, E9320 or N1920 sensor.
NOTE
SENS:FREQ:STAR, SENS:FREQ:STOP and SENS:FREQ:STEP are allowed to be set in
any desirable sequence.
Frequency step size calculated will be rounded to the nearest kHz with the
minimum size of 1 kHz. When frequency range is less than frequency sweep
step, the remaining steps will be repeated with the last frequency point.
Item Description Range of Values
numeric_value A numeric value for number of step in the average trigger
frequency sweep:
DEF: the default value is 0
MIN: 0
MAX: 2048
0 to 2048
DEF
MIN
MAX
SENS
SENS2
:
1
FREQ
:CW
:FIX
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space
:STEP