2 MEASurement Commands
150 N1911A/1912A P-Series Power Meters Programming Guide
Example
Error Messages
INITiate:CONTinuous must be set to OFF, otherwise error –213,
“INIT ignored” occurs.
–If TRIGger:SOURce is set to BUS or HOLD, error –214, “Trigger deadlock” occurs.
If the expected value and resolution parameters are not the same as the
current expected value and resolution settings on the specified window,
error –221, “Settings conflict” occurs.
READ1:REL? DEF,1,(@2) This command queries the upper window/
upper measurement relative measurement
of Channel B, using the current sensor
range and a resolution of 1.