MEASurement Commands 2
N1911A/1912A P-Series Power Meters Programming Guide 155
Example
Error Messages
INITiate:CONTinuous must be set to OFF on both channels, otherwise error
–213, “INIT ignored” occurs.
–If TRIGger:SOURce is set to BUS or HOLD on either channel, error –214, “Trigger
deadlock” occurs.
If the resolution parameter is not the same as the current resolution setting on
the specified window, error –221, “Settings conflict” occurs.
READ1:DIFF:REL? DEF,4,
(@2),(@1)
This command queries the upper
window/upper measurement
relative difference measurement of
Channel B -Channel A, using the
current sensor range and a
resolution setting of 4 on both
channels.