MEASurement Commands 2
N1911A/1912A P-Series Power Meters Programming Guide 165
[a] The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the command sub-systems. The parameters must
be entered in the specified order. If parameters are omitted, they default from the right. The parameter DEFault is used as a place holder.
Specifying DEF leaves the parameter value unchanged.
[b] When the measurement result is linear this parameter represents the number of significant digits. When the measurement result is
logarithmic 1 to 4 represents of 1, 0.1, 0.01 and 0.001 respectively.
Example
source list The channel which the command is implemented on.
If unspecified the current window setup is used. However, on the
N1912A, if the window shows a ratio or difference measurement,
the upper window defaults to Channel A and the lower window
to Channel B.
(@1)
(@2) (N1912A only)
MEAS1:REL? -50DBM,2,(@2) This command queries the upper window/
upper measurement relative measurement of
Channel B, using an expected power level of
–50 dBm and a resolution setting of 2.
Item Description/Default Range of Values