CALCulate Subsystem 3
N1911A/1912A P-Series Power Meters Programming Guide 203
Query
CALCulate[1]|2|3|4:MATH[:EXPRession]?
The query returns the current math measurement setting on the specified window.
Query Example
Error Messages
– For the single channel N1911A power meter: if <string> is not set to “(SENS1)”
while SENSe:SPEed is set to 200, error –221, “Settings Conflict” occurs.
– For the dual channel N1912A power meter: if <string> is not set to “(SENS1)”
or “(SENS2)” while SENS1:SPEEd or SENS2:SPEEd is set to 200, error –221,
“Settings Conflict” occurs.
CALC1:MATH? This command queries the current setting
of the math expression on the upper
window/upper measurement.