CALibration Subsystem 4
N1911A/1912A P-Series Power Meters Programming Guide 217
Syntax
Query Example
Error Messages
If the calibration was not carried out successfully the error –231, “Data
Questionable; CAL ERROR” occurs. If you are using an N1912A the error
message specifies which channel failed calibration.
If zeroing was not carried out successfully the error –231, “Data Questionable;
ZERO ERROR” occurs. If you are using an N1912A the error message specifies
which channel failed calibration.
If there is no sensor connected, the error –241, “Hardware Missing” occurs.
CAL1:ALL? This command causes the power meter to
perform a calibration sequence on Channel A
and return a result.
:ALL
?
CAL
1
2