8 OUTPut Subsystem
314 N1911A/1912A P-Series Power Meters Programming Guide
Query
OUTPut:TRIGger[:STATe]?
The query command enters a 1 or 0 into the output buffer indicating whether or
not the trigger output signal is enabled or disabled.
1 is returned when the trigger output signal is enabled
0 is returned when the trigger output signal is disabled
Query Example
Error Messages
If 8480, N8480, E4410, E9300 or E9320 sensor is connected and the trigger
source is not set to external, error –221 “Settings conflict” occurs.
OUTP:TRIG:STAT? This command queries the status of the
trigger output signal.