9 PSTatistic Subsystem
342 N1911A/1912A P-Series Power Meters Programming Guide
Reset Condition
On reset, the CCDF cummulative counts will be set to the default value, 100 M
samples.
Query
PSTatistic[1]|2:CCDF:COUNt?
The query returns the current numeric value of the CCDF cummulative count for
the respective channel selected.
Query Example
Error Messages
If no sensor or wrong sensor is connected to the channel, error –241
"Hardware missing" occurs.
If channel is not in free-run acquisition mode, error –221 "Settings conflict"
occurs.
If user inputs a count that is lesser than 100 M or greater than 10 G, error –222
"Data out of range" occurs. Any input count that is within range will be
normalized and round down to the nearest 100 M.
PST1:CCDF:COUN? This command queries the numeric value of
CCDF cummulative counts for Channel A.