9 PSTatistic Subsystem
342 N1911A/1912A P-Series Power Meters Programming Guide
Reset Condition
On reset, the CCDF cummulative counts will be set to the default value, 100 M
samples.
Query
PSTatistic[1]|2:CCDF:COUNt?
The query returns the current numeric value of the CCDF cummulative count for
the respective channel selected.
Query Example
Error Messages
– If no sensor or wrong sensor is connected to the channel, error –241
"Hardware missing" occurs.
– If channel is not in free-run acquisition mode, error –221 "Settings conflict"
occurs.
– If user inputs a count that is lesser than 100 M or greater than 10 G, error –222
"Data out of range" occurs. Any input count that is within range will be
normalized and round down to the nearest 100 M.
PST1:CCDF:COUN? This command queries the numeric value of
CCDF cummulative counts for Channel A.