10 SENSe Subsystem
384 N1911A/1912A P-Series Power Meters Programming Guide
Query Example
Error Messages
– If the command is used when a sensor other than a P-Series or E9320 power
sensor is connected, error –241, “Hardware missing” occurs
– If the command is used when an E9320 sensor is connected and set to
AVERage mode rather than NORMal mode, the error –221, “Settings Conflict”
occurs.
SENS2:AVER2? This command queries whether averaging is
on or off for Channel B.