SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 391
Example
Query
[SENSe[1]]|SENSe2:BUFFer:MTYP?
This query is used to retrieve measurement type settings.
Query Example
On Reset
On *RST, the value is set to AVER
Error Messages
– If no sensor or wrong sensor is connected to the channel, error –241,
“Hardware missing” occurs.
– If E9320 or N1920 sensor is connected but acquisition mode is in free run,
error –221, “Setting conflict. Invalid acquisition mode” occurs.
– If E9320 or N1920 sensor is connected but sensor mode is average, error
–221, “Setting conflict. Invalid acquisition mode” occurs.
– If E9320 or N1920 sensor is connected but trigger source is not external,
error-221 “Setting conflict. Invalid acquisition mode” occurs.
– If parameter set is a string but it is invalid, error-224 “Illegal parameter value”
occurs.
BUFF:MTYP AVER This command sets the measurement type to
average for Channel A.
BUFF:MTYP? This query returns the measurement type to
average for Channel A.