10 SENSe Subsystem
404 N1911A/1912A P-Series Power Meters Programming Guide
Query
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3[:INPut]
[:MAGNitude]? [MIN|MAX]
The query returns the current setting of the duty cycle or the values associated
with MIN and MAX.
Query Example
Error Messages
If a duty cycle value is entered using
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3 while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error –221, “Settings Conflict”
occurs. However, the duty cycle value is set but the
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3:STATe command is not
automatically set ON.
If this command is used when an E4412A/E4413A power sensor is connected,
the error –310, “System error;Dty Cyc may impair accuracy with ECP sensor”
occurs. If you are using a dual channel power meter the error message
specifies the channel.
If this command is used when a E9320 Series power sensor is set to NORMal
mode, the error –221, “Settings Conflict” occurs.
If this command is used when a P-Series power sensor is connected, the error
–241, “Hardware missing” occurs.
CORR:GAIN3? This command queries the current setting of
the duty cycle for Channel A.