SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 413
Query Example
Error Messages
If a loss/gain correction value is entered using
[SENSe[1]]|SENSe2:CORRection:GAIN2[:INPut][:MAGNitude] while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error –221, “Settings Conflict”
occurs. However, the correction value is set but the
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe command is not automatically
set ON.
–The SENSe[1]]|SENSe2:CORRection:GAIN2[:INPut][:MAGNitude] command
can be used for the 8480 Series power sensor when no sensor calibration table
has been set up.
CORR:GAIN2? This command queries the current setting of
the channel offset on Channel A.