1 Power Meter Remote Operation
42 N1911A/1912A P-Series Power Meters Programming Guide
Example 2 - Specifying the Source List Parameter
The CONFigure and READ? commands have three optional parameters, an
expected power value, a resolution and a source list. These parameters must be
entered in the specified order. If parameters are omitted, they default from the
right. The parameter DEFault is used as a place holder.
The following examples use the source list parameter to specify the measurement
channel as Channel A. The expected power and resolution parameters are
defaulted, leaving them at their current settings. The measurement is carried out
on the upper window.
Although the READ? and FETCh? queries have three optional parameters it is not
necessary to define them as shown in these examples. If they are defined they
must be identical to those defined in the CONFigure command otherwise an error
occurs.
Using READ?
NOTE
For the N1911A it is not necessary to specify a channel as only one channel is
available.
ABOR1 Aborts Channel A
CONF1 DEF,DEF,(@1) Configures the upper window to make a
Channel A measurement using the current
expected power and resolution settings
READ1? Takes the upper window’s measurement