10 SENSe Subsystem
428 N1911A/1912A P-Series Power Meters Programming Guide
Syntax
Parameters
[a] When a channel is set to NORMal or DOUBle, TRIG:COUNt is set automatically to 1.
Example
Reset Condition
On reset, the speed is set to NORMal.
Query
[SENSe[1]]|SENSe2:MRAT?
The query returns the current speed setting, either NORMal, DOUBle or FAST.
Item Description/Default Range of Values
character_data A numeric value for the measurement speed:
NORMal: 20 readings/second
DOUBle: 40 readings/second
FAST: up to 1000 readings/second
The default is NORMal.
NORMal
[a]
DOUBle
[a]
FAST
MRAT DOUBle This command sets the Channel A speed to
40 readings/second.
SENS
1
SENS2
:
Space
?
numeric_value
MRAT