Reset Condition
On reset, autoranging is enabled.
Query
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO?
The query enters a 1 or 0 into the output buffer indicating the status of
autoranging.
1 is returned when autoranging is enabled
0 is returned when autoranging is disabled
Query Example
Error Messages
If this command is set to OFF when there is not an E-Series power sensor or
N8480 Series power sensor (excluding Option CFT) connected, the error –241,
“Hardware missing” occurs.
If this command is set to ON when E9320 power sensor is in Normal and
Triggered modes, the error -221, “Setting conflicts” occurs.
POW:AC:RANG:AUTO? This command queries whether auto ranging