1 Power Meter Remote Operation
44 N1911A/1912A P-Series Power Meters Programming Guide
Using READ?
Some fine tuning of measurements can be performed using the CONFigure and
READ? commands. For example, in the above program segment some fine tuning
can be performed by setting the filter length to 1024 and the trigger delay off.
1 ABOR2
2 CONF1 -50,DEF,(@2)
3 SENS2:AVER:COUN 1024
4 TRIG2:DEL:AUTO OFF
5 READ1?
Using INITiate and FETCh?
Some fine tuning of measurements can be carried out using the CONFigure
command and INITiate and FETCh? commands. For example, in the above
program segment some fine tuning can be carried out by setting the filter length
to 1024 and the trigger delay off.
ABOR2 Aborts Channel B
CONF1 -50,DEF,(@2) Configures the upper window to make a
Channel B measurement using an expected
power of –50 dBm and the current resolution
setting
READ1? Takes the upper window’s measurement
ABOR2 Aborts Channel B
CONF1 -50,DEF,(@2) Configures the upper window to make a
Channel B measurement using an expected
power of –50 dBm and the current resolution
setting
INIT2 Causes Channel B to make a measurement
FETC1? -50,DEF,(@2) Retrieves the upper window’s measurement