SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 455
Example
Reset Condition
On reset, X Start = 0 s, X Scale = 10 ms, Y Max = 20 dBm, Y Scale = 7 dB.
Error Messages
– If no sensor or wrong sensor is connected to the channel, error –241
"Hardware missing" occurs.
– If channel is not in trigger acquisition mode, error –221 "Settings conflict"
occurs.
SENS:TRAC:AUT This command triggers the Auto Scaling for
Channel A.
SENS2:TRAC:AUT This command triggers the Auto Scaling for
Channel B.