EasyManua.ls Logo

Keysight N1911A

Keysight N1911A
771 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 461
Reset Condition
On reset, X Start = 0 s; X Scale = 10 µs.
Query
SENSe[1]|2:TRACe:X:SCALe:PDIV?
The query returns the current scale setting of X-axis in numerical value.
Query Example
Error Messages
If no sensor or wrong sensor is connected to the channel, error –241
"Hardware missing" occurs.
If channel is not in trigger acquisition mode, error –221 "Settings conflict"
occurs.
If limits of the values keyed in are exceeded, error –222 "Data out of range;
upper (or lower) limit exceeded; no change" occurs.
SENS:TRAC:X:SCAL:PDIV? This command queries the Channel A current
X-axis scale setting in numerical value.

Table of Contents

Other manuals for Keysight N1911A

Related product manuals